Are
you having problem with your GX-CT01? Does your unit displays
inaccurate readings or does it show 0.00 every time you make
measurements? Here are some tips on how to resolve your problem
without any hassle on contacting your seller.
If
your unit is giving inaccurate readings and every time you measure
the unit is giving different measurement at the same spot. Well you
must remember that error has factors to consider, 2 of the major
factors are machine error and human error. Let's discuss further in
this blog what does it mean.
The
unit is giving different readings because, in every press you make,
you're exerting different pressure on the item, does caused to give
different reading, you also need to consider the accuracy range of
the unit 2%RDG±0.02mm, if it is within the accuracy range then your
item is working just fine. If the unit is giving way more different
reading than the correct measurement, I think your unit needs a zero
calibration. Just follow the calibration process I indicated below,
you must remember to clean the sensor of the item first with soft
cloth to ensure the accuracy of the item. Also, make sure that the
unit of measurement corresponds to the material you're going to test
("NFe" for non-iron and "Fe" for iron base).
First test the unit to the iron and non iron base included in the
set.
Zero Calibration Process:
1.
Change the battery of the item with brand new. The low voltage
battery can cause inaccuracy to measurement. And turn on the unit.
2.
Press to the corresponding base you're going to calibrate with.
3.
Long press the zero calibration button, then release the button when
the screen displays CAL. Don't leave the base not until the unit
displays 0.00.
4.
Once 0.00 displays, it means the unit is successfully calibrated.
After calibrating the unit to both bases, you may try to test the
unit.
Your
unit is always displaying and giving a measurement of 0.00 to each
and every measurement you make? Don't think that the item is
defective, perhaps, it is only because the value of the material
you're measuring has a lower value than of the base included in the
set. If the unit keeps displaying 0.00, you may try to calibrate the
meter on a different material which doesn't have coating.
For
example, the value of the car you will test is 4 microns. Then you
may calibrate the meter with a non coating material with a value of 4
microns. After calibration please test the unit.
If
after following all my suggestion to this page and the problem wasn't
resolved, I suggest that you contact your seller for further
assistance. Does this page answer your questions? If you have further
question regarding your GX-CT01, you may comment your questions or
concerns, we will be happy to answer you.